RI 6026 Calibration Study Of A Peak-Reading Accelerograph

- Organization:
- The National Institute for Occupational Safety and Health (NIOSH)
- Pages:
- 10
- File Size:
- 2073 KB
- Publication Date:
- Jan 1, 1962
Abstract
A Blastcorder4 was calibrated on the Bureau of Mines shaking table in the frequency range of 12 to 40 c.p.s. and in the amplitude range' of 0.25 to 1.0 g. The results of this calibration study show that in the frequency range of 12 to 30 c.p.s. the average accuracy of the Blastcorder is approximately ± 0.1 g.