RI 3077 Note On Copper Constantan Thermocouple Calibration Below O°C. ? Introduction

The National Institute for Occupational Safety and Health (NIOSH)
R. Wiebe
Organization:
The National Institute for Occupational Safety and Health (NIOSH)
Pages:
9
File Size:
2533 KB
Publication Date:
Jan 1, 1931

Abstract

During a recent investigation the authors were interested in determining: (1) How many points were necessary to locate the deviation curve for copper-constantan thermocouples within an accuracy of about 0.05° C., used in connection with the, "standard" thermocouple of L. H. Adams (2) Whether copper-constantan thermocouples showing considerable deviations from the "standard" were consistent throughout their calibration. (3) How reliable the extrapolated portion is - that is, the interval between -183 and -220° C. of the "standard" thermocouple.