IC 8754 Noise Dosimeter Performance - A Second Evaluation

The National Institute for Occupational Safety and Health (NIOSH)
Timothy Y. Yen
Organization:
The National Institute for Occupational Safety and Health (NIOSH)
Pages:
43
File Size:
15853 KB
Publication Date:
Jan 1, 1977

Abstract

This Bureau of Mines report evaluates audio dosimeters, 3 sample units from each of 10 currently marketed brands. Each dosimeter was treated as a "black box" with the acoustic stimulus as the input and the accumulated dosage as the output. The dosimeter microphones were exposed to a sound field gener¬ated in an anechoic chamber. Characteristics evaluated include frequency response (100 to 8,000 Hz), level-time trade-off rate (88 to 117 db), status of calibration, crest-handling capability, and power consumption rate. Experi¬mental errors revealed by replication of the tests were found to be generally much smaller than variations in performance among the samples of any brand, and possibilities of further reducing the magnitude of these errors are discussed.