Abstract
This paper examines the viability of using the QEM*SEM automated scanning electron microscope for the determination iron sulfide and calcite volumetric percentages. The analytical reliability was verified by comparing elemental abundances calculated for fourteen samples using the QEM*SEM volumetric percentages and appropriate mineral compositions and elemental abundances determined by standard laboratory procedures. The abundances were comparable, thus demonstrating suitability of this technique for evaluation of the acid neutralization and generation potentials of mining wastes.
Citation
APA: